Transfer Shortcuts¶
The Transfer command is a nice way to wrap up the most common
liquid-handling actions we take. Instead of having to write
loop
and
if
statements, we can simply use the
transfer()
command, making Python protocol both easier to write and read!
Transfer¶
Most of time, a protocol is really just looping over some
wells, aspirating, and then dispensing. Even though they are
simple in nature, these loops take up a lot of space. The
pipette.transfer()
command takes care of those common loops. It will combine
aspirates and dispenses automatically, making your protocol
easier to read and edit.
'''
Examples in this section expect the following
'''
from opentrons import containers, instruments
plate = containers.load('96-flat', 'B1')
tiprack = containers.load('tiprack-200ul', 'A1')
trash = containers.load('point', 'D2')
pipette = instruments.Pipette(
axis='b',
max_volume=200,
tip_racks=[tiprack],
trash_container=trash)
Basic¶
The example below will transfer 100 uL from well
'A1'
to well
'B1', automatically picking up a new tip and then dropping it
when finished.
pipette.transfer(100, plate.wells('A1'), plate.wells('B1'))
Transfer commands will automatically create entire series of
aspirate(),
dispense(), and other
Pipette
commands. We can print out all commands to see what it did
in the previous example:
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B1>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
Large Volumes¶
Volumes larger than the pipette’s
max_volume
will automatically divide into smaller transfers.
pipette.transfer(700, plate.wells('A2'), plate.wells('B2'))
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 200.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 200.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Aspirating 200.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 200.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Aspirating 150.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 150.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Aspirating 150.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 150.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
Multiple Wells¶
Transfer commands are most useful when moving liquid between multiple wells.
pipette.transfer(100, plate.cols('A'), plate.cols('B'))
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A3>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B3>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A4>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B4>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A5>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B5>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A6>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B6>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A7>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B7>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A8>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B8>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A9>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B9>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A10>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B10>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A11>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B11>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A12>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B12>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
One to Many¶
You can transfer from a single source to multiple destinations, and the other way around (many sources to one destination).
pipette.transfer(100, plate.wells('A1'), plate.rows('2'))
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well C2>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well D2>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well E2>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well F2>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well G2>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well H2>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
Few to Many¶
What happens if, for example, you tell your pipette to transfer from 4 source wells to 2 destination wells? The transfer command will attempt to divide the wells evenly, or raise an error if the number of wells aren’t divisible.
pipette.transfer(
100,
plate.wells('A1', 'A2', 'A3', 'A4'),
plate.wells('B1', 'B2'))
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A3>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A4>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
List of Volumes¶
Instead of applying a single volume amount to all source/destination wells, you can instead pass a list of volumes.
pipette.transfer(
[20, 40, 60],
plate.wells('A1'),
plate.wells('B1', 'B2', 'B3'))
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 20.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 20.0 at <Deck><Slot B1><Container 96-flat><Well B1>
Aspirating 40.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 40.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Aspirating 60.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 60.0 at <Deck><Slot B1><Container 96-flat><Well B3>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
Volume Gradient¶
Create a linear gradient between a start and ending volume (uL). The start and ending volumes must be the first and second elements of a tuple.
pipette.transfer(
(100, 30),
plate.wells('A1'),
plate.rows('2'))
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Aspirating 90.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 90.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Aspirating 80.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 80.0 at <Deck><Slot B1><Container 96-flat><Well C2>
Aspirating 70.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 70.0 at <Deck><Slot B1><Container 96-flat><Well D2>
Aspirating 60.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 60.0 at <Deck><Slot B1><Container 96-flat><Well E2>
Aspirating 50.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 50.0 at <Deck><Slot B1><Container 96-flat><Well F2>
Aspirating 40.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 40.0 at <Deck><Slot B1><Container 96-flat><Well G2>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well H2>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
Distribute and Consolidate¶
Save time and tips with the
distribute()
and
consolidate()
commands. These are nearly identical to
transfer(), except that they will combine multiple transfer’s
into a single tip.
'''
Examples in this section expect the following
'''
from opentrons import containers, instruments
plate = containers.load('96-flat', 'B1')
tiprack = containers.load('tiprack-200ul', 'A1')
trash = containers.load('point', 'D2')
pipette = instruments.Pipette(
axis='b',
max_volume=200,
tip_racks=[tiprack],
trash_container=trash)
Consolidate¶
Volumes going to the same destination well are combined within the same tip, so that multiple aspirates can be combined to a single dispense.
pipette.consolidate(30, plate.rows('2'), plate.wells('A1'))
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well C2>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well D2>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well E2>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well F2>
Dispensing 180.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well G2>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well H2>
Dispensing 60.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
If there are multiple destination wells, the pipette will never combine their volumes into the same tip.
pipette.consolidate(30, plate.rows('2'), plate.wells('A1', 'A2'))
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well C2>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well D2>
Dispensing 120.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well E2>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well F2>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well G2>
Aspirating 30.0 at <Deck><Slot B1><Container 96-flat><Well H2>
Dispensing 120.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
Distribute¶
Volumes from the same source well are combined within the same tip, so that one aspirate can provide for multiple dispenses.
pipette.distribute(55, plate.wells('A1'), plate.rows('2'))
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 165.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 55.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 55.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Dispensing 55.0 at <Deck><Slot B1><Container 96-flat><Well C2>
Aspirating 165.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 55.0 at <Deck><Slot B1><Container 96-flat><Well D2>
Dispensing 55.0 at <Deck><Slot B1><Container 96-flat><Well E2>
Dispensing 55.0 at <Deck><Slot B1><Container 96-flat><Well F2>
Aspirating 110.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 55.0 at <Deck><Slot B1><Container 96-flat><Well G2>
Dispensing 55.0 at <Deck><Slot B1><Container 96-flat><Well H2>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
If there are multiple source wells, the pipette will never combine their volumes into the same tip.
pipette.distribute(30, plate.wells('A1', 'A2'), plate.rows('2'))
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 120.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well C2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well D2>
Aspirating 120.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well E2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well F2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well G2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well H2>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
Disposal Volume¶
When dispensing multiple times from the same tip, it is
recommended to aspirate an extra amount of liquid to be
disposed of after distributing. This added
disposal_vol
can be set as an optional argument.
pipette.distribute(
30,
plate.wells('A1', 'A2'),
plate.rows('2'),
disposal_vol=10) # include extra liquid to make dispenses more accurate
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 130.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well C2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well D2>
Blowing out at <Deck><Slot D2><Container point><Well A1>
Aspirating 130.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well E2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well F2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well G2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well H2>
Blowing out at <Deck><Slot D2><Container point><Well A1>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
Note
If you do not specify a
disposal_vol, the pipette will by default use a
disposal_vol
equal to it’s
min_volume. This tutorial has not given the pipette any
min_volume, so below is an example of allowing the pipette’s
min_volume
to be used as a default for
disposal_vol.
pipette.min_volume = 20 # `min_volume` is used as default to `disposal_vol`
pipette.distribute(
30,
plate.wells('A1', 'A2'),
plate.rows('2'))
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 140.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well C2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well D2>
Blowing out at <Deck><Slot D2><Container point><Well A1>
Aspirating 140.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well E2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well F2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well G2>
Dispensing 30.0 at <Deck><Slot B1><Container 96-flat><Well H2>
Blowing out at <Deck><Slot D2><Container point><Well A1>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
Transfer Options¶
There are other options for customizing your transfer command:
'''
Examples in this section expect the following
'''
from opentrons import containers, instruments
plate = containers.load('96-flat', 'B1')
tiprack = containers.load('tiprack-200ul', 'A1')
trash = containers.load('point', 'D2')
pipette = instruments.Pipette(
axis='b',
max_volume=200,
tip_racks=[tiprack],
trash_container=trash)
Always Get a New Tip¶
Transfer commands will by default use the same one tip for each well, then finally drop it in the trash once finished.
The pipette can optionally get a new tip at the beginning of each aspirate, to help avoid cross contamination.
pipette.transfer(
100,
plate.wells('A1', 'A2', 'A3'),
plate.wells('B1', 'B2', 'B3'),
new_tip='always') # always pick up a new tip
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B1>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well B1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well C1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A3>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B3>
Drop_tip at <Deck><Slot D2><Container point><Well A1>
Never Get a New Tip¶
For scenarios where you instead are calling
pick_up_tip()
and
drop_tip()
elsewhere in your protocol, the transfer command can ignore
picking up or dropping tips.
pipette.transfer(
100,
plate.wells('A1', 'A2', 'A3'),
plate.wells('B1', 'B2', 'B3'),
new_tip='never') # never pick up or drop a tip
for c in robot.commands():
print(c)
will print out...
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B2>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A3>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B3>
Trash or Return Tip¶
By default, the transfer command will drop the
pipette’s tips in the trash container. However, if you
wish to instead return the tip to it’s tip rack, you
can set
trash=False.
pipette.transfer(
100,
plate.wells('A1'),
plate.wells('B1'),
trash=False) # do not trash tip
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well B1>
Returning tip
Drop_tip at <Deck><Slot A1><Container tiprack-200ul><Well A1>
Touch Tip¶
A touch-tip can be performed after every aspirate and
dispense by setting
touch_tip=True.
pipette.transfer(
100,
plate.wells('A1'),
plate.wells('A2'),
touch_tip=True) # touch tip to each well's edge
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Touching tip
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Touching tip
Drop_tip at <Deck><Slot D2><Container point><Well A1>
Blow Out¶
A blow-out can be performed after every dispense that leaves
the tip empty by setting
blow_out=True.
pipette.transfer(
100,
plate.wells('A1'),
plate.wells('A2'),
blow_out=True) # blow out droplets when tip is empty
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Blowing out
Drop_tip at <Deck><Slot D2><Container point><Well A1>
Mix Before/After¶
A mix can be performed before every aspirate by setting
mix_before=. The value of
mix_before=
must be a tuple, the 1st value is the number of repetitions,
the 2nd value is the amount of liquid to mix.
pipette.transfer(
100,
plate.wells('A1'),
plate.wells('A2'),
mix_before=(2, 50), # mix 2 times with 50uL before aspirating
mix_after=(3, 75)) # mix 3 times with 75uL after dispensing
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Mixing 2 times with a volume of 50ul
Aspirating 50 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 50
Aspirating 50
Dispensing 50
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Mixing 3 times with a volume of 75ul
Aspirating 75 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 75.0
Aspirating 75
Dispensing 75.0
Aspirating 75
Dispensing 75.0
Drop_tip at <Deck><Slot D2><Container point><Well A1>
Air Gap¶
An air gap can be performed after every aspirate by setting
air_gap=int, where the value is the volume of air in microliters to
aspirate after aspirating the liquid.
pipette.transfer(
100,
plate.wells('A1'),
plate.wells('A2'),
air_gap=20) # add 20uL of air after each aspirate
for c in robot.commands():
print(c)
will print out...
Picking up tip from <Deck><Slot A1><Container tiprack-200ul><Well A1>
Aspirating 100.0 at <Deck><Slot B1><Container 96-flat><Well A1>
Air gap
Aspirating 20
Dispensing 20 at <Deck><Slot B1><Container 96-flat><Well A2>
Dispensing 100.0 at <Deck><Slot B1><Container 96-flat><Well A2>
Drop_tip at <Deck><Slot D2><Container point><Well A1>