Liquid Handling¶
The
instruments
module gives your protocol access to the
Pipette, which is what you will be primarily using to create protocol
commands.
Creating a Pipette¶
'''
Examples in this section require the following
'''
from opentrons import instruments
Axis and Max Volume¶
To create a
Pipette, you must give it an axis and a max_volume. The axis can
be either
'a'
or
'b', and the volume is whatever your hand pipette is
calibrated for. In this example, we are using a 200uL
pipette.
pipette = instruments.Pipette(
axis='b',
name='my-p200',
max_volume=200)
Minimum Volume¶
The minimum allowed volume can be set for each pipette. If your protocol attempts to aspirate or dispense a volume below this volume, the API will give you a warning.
pipette = instruments.Pipette(
axis='b',
name='my-p200',
max_volume=200,
min_volume=20)
Channels¶
Pipettes can also be assigned a number of channels, either
channel=1
or
channel=8. If you do not specify, it will default to
channel=1
channel.
pipette = instruments.Pipette(
axis='b',
name='my-p200-multichannel',
max_volume=200,
min_volume=20,
channels=8)
Plunger Speeds¶
The speeds at which the pipette will aspirate and dispense
can be set through
aspirate_speed
and
dispense_speed. The values are in millimeters/minute, and default to
aspirate_speed=300
and
dispense_speed=500.
pipeipette = instruments.Pipette(
axis='b',
name='my-p200-multichannel',
max_volume=200,
min_volume=20,
channels=8,
aspirate_speed=200,
dispense_speed=600)
Tip Handling¶
When we handle liquids with a pipette, we are constantly exchanging old, used tips for new ones to prevent cross-contamination between our wells. To help with this constant need, we describe in this section a few methods for getting new tips, and removing tips from a pipette.
This section demonstrates the options available for controlling tips
'''
Examples in this section expect the following
'''
from opentrons import containers, instruments
trash = containers.load('point', 'D2')
tiprack = containers.load('tiprack-200ul', 'B1')
pipette = instruments.Pipette(axis='a')
Pick Up Tip¶
Before any liquid handling can be done, your pipette must
have a tip on it. The command
pick_up_tip()
will move the pipette over to the specified tip, the press
down into it to create a vacuum seal. The below example
picks up the tip at location
'A1'.
pipette.pick_up_tip(tiprack.wells('A1'))
Drop Tip¶
Once finished with a tip, the pipette will autonomously
remove the tip when we call
drop_tip(). We can specify where to drop the tip by passing in a
location. The below example drops the tip back at its
originating location on the tip rack.
pipette.drop_tip(tiprack.wells('A1'))
Instead of returning a tip to the tip rack, we can also drop it in a trash container.
pipette.pick_up_tip(tiprack.wells('A2'))
pipette.drop_tip(trash)
Return Tip¶
When we need to return the tip to its originating location
on the tip rack, we can simply call
return_tip(). The example below will automatically return the tip to
'A3'
on the tip rack.
pipette.pick_up_tip(tiprack.wells('A3'))
pipette.return_tip()
Tips Iterating¶
Automatically iterate through tips and drop tip in trash by attaching containers to a pipette
'''
Examples in this section expect the following
'''
from opentrons import containers, instruments
trash = containers.load('point', 'D2')
tip_rack_1 = containers.load('tiprack-200ul', 'B1')
tip_rack_2 = containers.load('tiprack-200ul', 'B2')
Attach Tip Rack to Pipette¶
Tip racks and trash containers can be “attached” to a pipette when the pipette is created. This give the pipette the ability to automatically iterate through tips, and to automatically send the tip to the trash container.
Trash containers can be attached with the option
trash_container=TRASH_CONTAINER.
Multiple tip racks are can be attached with the option
tip_racks=[RACK_1,
RACK_2,
etc...
].
pipette = instruments.Pipette(
axis='b',
tip_racks=[tip_rack_1, tip_rack_2],
trash_container=trash
)
Note
The
tip_racks=
option expects us to give it a Python list, containing
each tip rack we want to attach. If we are only
attaching one tip rack, then the list will have a length
of one, like the following:
tip_racks=[tiprack]
Iterating Through Tips¶
Now that we have two tip racks attached to the pipette, we
can automatically step through each tip whenever we call
pick_up_tip(). We then have the option to either
return_tip()
to the tip rack, or we can
drop_tip()
to remove the tip in the attached trash container.
pipette.pick_up_tip() # picks up tip_rack_1:A1
pipette.return_tip()
pipette.pick_up_tip() # picks up tip_rack_1:A2
pipette.drop_tip() # automatically drops in trash
# use loop to pick up tips tip_rack_1:A3 through tip_rack_2:H12
for i in range(94 + 96):
pipette.pick_up_tip()
pipette.return_tip()
If we try to
pick_up_tip()
again when all the tips have been used, the Opentrons API
will show you an error.
Note
If you run the cell above, and then uncomment and run the cell below, you will get an error because the pipette is out of tips.
# this will raise an exception if run after the previous code block
# pipette.pick_up_tip()
Select Starting Tip¶
Calls to
pick_up_tip()
will by default start at the attached tip rack’s
'A1'
location. If you however want to start automatic tip
iterating at a different tip, you can use
start_at_tip().
pipette.reset()
pipette.start_at_tip(tip_rack_1['C3'])
pipette.pick_up_tip() # pick up C3 from "tip_rack_1"
pipette.return_tip()
Get Current Tip¶
Get the source location of the pipette’s current tip
by calling
current_tip(). If the tip was from the
'A1'
position on our tip rack,
current_tip()
will return that position.
print(pipette.current_tip()) # is holding no tip
pipette.pick_up_tip()
print(pipette.current_tip()) # is holding the next available tip
pipette.return_tip()
print(pipette.current_tip()) # is holding no tip
will print out...
None
<Well D3>
None
Liquid Control¶
This is the fun section, where we get to move things around
and pipette! This section describes the
Pipette
object’s many liquid-handling commands, as well as how
to move the
robot.
'''
Examples in this section expect the following
'''
from opentrons import containers, instruments
plate = containers.load('96-flat', 'B1')
pipette = instruments.Pipette(axis='b', max_volume=200)
Aspirate¶
To aspirate is to pull liquid up into the pipette’s tip. When calling aspirate on a pipette, we can specify how many micoliters, and at which location, to draw liquid from:
pipette.aspirate(50, plate.wells('A1')) # aspirate 50uL from plate:A1
Now our pipette’s tip is holding 50uL.
We can also simply specify how many microliters to aspirate,
and not mention a location. The pipette in this circumstance
will aspirate from it’s current location (which we
previously set as
plate.wells('A1')).
pipette.aspirate(50) # aspirate 50uL from current position
Now our pipette’s tip is holding 100uL.
We can also specify only the location to aspirate from. If we do not tell the pipette how many micoliters to aspirate, it will by default fill up the remaining volume in it’s tip. In this example, since we already have 100uL in the tip, the pipette will aspirate another 100uL
pipette.aspirate(plate.wells('A2')) # aspirate until pipette fills from plate:A2
Dispense¶
To dispense is to push out liquid from the pipette’s
tip. It’s usage in the Opentrons API is nearly
identical to
aspirate(), in that you can specify microliters and location, only
microliters, or only a location:
pipette.dispense(50, plate.wells('B1')) # dispense 50uL to plate:B1
pipette.dispense(50) # dispense 50uL to current position
pipette.dispense(plate.wells('B2')) # dispense until pipette empties to plate:B2
That final dispense without specifying a micoliter amount
will dispense all remaining liquids in the tip to
plate.wells('B2'), and now our pipette is empty.
Blow Out¶
To blow out is to push an extra amount of air through the pipette’s tip, so as to make sure that any remaining droplets are expelled.
When calling
blow_out()
on a pipette, we have the option to specify a location to
blow out the remaining liquid. If no location is specified,
the pipette will blow out from it’s current position.
pipette.blow_out() # blow out over current location
pipette.blow_out(plate.wells('B3')) # blow out over current plate:B3
Touch Tip¶
To touch tip is to move the pipette’s currently attached tip to the edges of a well, for the purpose of knocking off any droplets that might be hanging from the tip.
When calling
touch_tip()
on a pipette, we have the option to specify a location where
the tip will touch the inner walls. If no location is
specified, the pipette will touch tip inside it’s
current location.
pipette.touch_tip() # touch tip within current location
pipette.touch_tip(-2) # touch tip 2mm below the top of the current location
pipette.touch_tip(plate.wells('B1')) # touch tip within plate:B1
Mix¶
Mixing is simply performing a series of
aspirate()
and
dispense()
commands in a row on a single location. However, instead of
having to write those commands out every time, the Opentrons
API allows you to simply say
mix().
The mix command takes three arguments:
mix(repetitions,
volume,
location)
pipette.mix(4, 100, plate.wells('A2')) # mix 4 times, 100uL, in plate:A2
pipette.mix(3, 50) # mix 3 times, 50uL, in current location
pipette.mix(2) # mix 2 times, pipette's max volume, in current location
Air Gap¶
Some liquids need an extra amount of air in the
pipette’s tip to prevent it from sliding out. A call
to
air_gap()
with a microliter amount will aspirate that much air into
the tip.
pipette.aspirate(100, plate.wells('B4'))
pipette.air_gap(20)
Moving¶
Demonstrates the different ways to control the movement of the Opentrons liquid handler during a protocol run.
'''
Examples in this section expect the following
'''
from opentrons import containers, instruments, robot
tiprack = containers.load('tiprack-200ul', 'A1')
plate = containers.load('96-flat', 'B1')
pipette = instruments.Pipette(axis='b')
Move To¶
Pipette’s are able to
move_to()
any location on the deck.
For example, we can move to the first tip in our tip rack:
pipette.move_to(tiprack.wells('A1'))
You can also specify at what height you would like the robot
to move to inside of a location using
top()
and
bottom()
methods on that location.
pipette.move_to(plate.wells('A1').bottom()) # move to the bottom of well A1
pipette.move_to(plate.wells('A1').top()) # move to the top of well A1
pipette.move_to(plate.wells('A1').bottom(2)) # move to 2mm above the bottom of well A1
pipette.move_to(plate.wells('A1').top(-2)) # move to 2mm below the top of well A1
The above commands will cause the robot’s head to
first move upwards, then over to above the target location,
then finally downwards until the target location is reached.
If instead you would like the robot to mive in a straight
line to the target location, you can set the movement
strategy to
'direct'.
pipette.move_to(plate.wells('A1'), strategy='direct')
Note
Moving with
strategy='direct'
will run the risk of colliding with things on your deck.
Be very careful when using the option.
Usually the
strategy='direct'
option is useful when moving inside of a well. Take a look
at the below sequence of movements, which first move the
head to a well, and use ‘direct’ movements
inside that well, then finally move on to a different well.
pipette.move_to(plate.wells('A1'))
pipette.move_to(plate.wells('A1').bottom(1), strategy='direct')
pipette.move_to(plate.wells('A1').top(-2), strategy='direct')
pipette.move_to(plate.wells('A1'))
Delay¶
To have your protocol pause for any given number of minutes
or seconds, simply call
delay()
on your pipette. The value passed into
delay()
is the number of minutes or seconds the robot will wait
until moving on to the next commands.
pipette.delay(seconds=2) # pause for 2 seconds
pipette.delay(minutes=5) # pause for 5 minutes
pipette.delay(minutes=5, seconds=2) # pause for 5 minutes and 2 seconds