Product Specifications
Included Parts¶
(1) Absorbance Plate Reader Module
(1) Caddy for deck mounting
(1) USB cable
(2) Spare deck slot screws (M4x10 mm socket head)
(1) Cleaning cloth
(1) Test certificate
Physical Specifications¶
| Specification | Description |
|---|---|
| Module dimensions | 155.3 mm L x 95.5 mm W x 57 mm H |
| Module weight | 790 g |
| Composition | Aluminum |
| Pollution degree | 2 |
| Service life | 10 years with average use of 4 hours per day |
Measurement Specifications¶
| Measurement method | Absorbance |
|---|---|
| Measurement techniques | Endpoint and kinetic |
| Detection | 96 photodiodes |
| Measurement range | 0–4.0 optical density (OD) |
| Resolution | 0.001 OD |
| Accuracy |
The maximum deviation between the determined value and the true value. At 405 nm:
At or above 450 nm:
|
| Reproducibility |
The maximum deviation between the determined values when the measurement is repeated directly.
|
| Linearity |
The maximum deviation between the true and the determined increase of the value. At 405 nm:
At or above 450 nm:
|
Detection Wavelengths¶
The Absorbance Plate Reader emits light in the visible spectrum at 450 nm (blue), 562 nm (green), 600 nm (orange), and 650 nm (red).
Status Light¶
The Absorbance Plate Reader has a single status light on the lid. It illuminates in different colors and patterns to indicate various operating conditions.
| Color | Pattern | Status |
|---|---|---|
| White | Solid | The module is on and ready. |
| Pulsing | Self-test after connecting to power. | |
| Various | Solid | Initialization/measurement in progress. The status light color corresponds to the selected light wavelength used for analysis. |
| Yellow | Pulsing | A well plate is inside the reader. |
| Red | Blinking | An error has occurred. |
Input and Output Connections¶
The Absorbance Plate Reader has the following power input requirements, which are met by its USB connection to a Flex robot.
- Input: Power from a USB port with 5 VDC and a maximum of 3 A.
- Power consumption: 2.5 W
- Fuse: 1 A (very fast acting)