Labware Definitions

Every piece of labware you use on an OT-2 requires a labware definition. Each definition contains all the information your robot needs to work with a piece of labware. This includes information about the physical shape of the labware, what pipettes can interact with it, and what the labware should be called in the Opentrons App. The OT-2 robot software and the Opentrons App include the labware definitions for everything available in the Opentrons Labware Library.

Custom labware is labware that's not listed in the Opentrons Labware Library. You can use other common or unique labware items with the OT-2 by accurately measuring and recording the characteristics of that object and saving that data in a JSON file. When imported into the app, the OT-2 and the Python Protocol API use that JSON data to interact with your labware. Opentrons provides tools and services, which we'll examine below, to help you use the OT-2 with custom labware.

Note

While you cannot create custom labware within the Python Protocol API, you can use custom labware with the available API methods. However, you need to define your custom labware first and import it into the Opentrons App. Then your custom labware is available to the Python API and the robot.

Custom Labware Creator

The Custom Labware Creator is a no-code, web-based tool that uses a graphical interface to help you create a labware definition file. Labware Creator produces a JSON labware definition file that you import into the Opentrons App. After that, your custom labware is available to the OT-2 and the Python API.

You can use the Custom Labware Creator if your labware meets the following criteria:

  • Wells and tubes are uniform and identical.
  • All rows are evenly spaced (the space between rows is equal).
  • All columns are evenly spaced (the space between columns is equal).
  • The labware fits perfectly in one deck slot.
Layout Description
Labware with 3 evenly spaced rows and 4 evenly spaced columns. Regular
All columns are evenly spaced and all rows are evenly spaced.
Columns do not need to have the same spacing as rows.
Labware with 3 evenly spaced rows and 4 evenly spaced columns on the left side of the labware. Regular
The grid does not have to be in the center of labware.

For other labware, consider the Custom Labware Service, outlined below. Or you can reference the complete JSON schema to create a labware definition from scratch, although this is not recommended.

Custom Labware Service

Get in touch with us if the labware you'd like to use isn't available in the library, if you can't create your own definitions, or if a custom item includes different shapes, sizes, or other irregularities described below.

  • Well or tube shapes vary.
  • Rows are not evenly spaced.
  • Columns are not evenly spaced.
  • The labware is smaller than one deck slot (requires adapter) or spans multiple deck slots.
Layout Description
Labware with 4 columns of 3 rows, separated into two groups. Irregular
Rows are evenly spaced but columns are not evenly spaced.
Labware with 3 square wells and 9 circular wells. Irregular
Columns/rows are evenly spaced but wells are not identical.
Labware with a 4-by-5 grid of wells and another 2-by-3 grid of wells. Irregular
There is more than one grid.

If you need help creating custom labware definitions, contact Opentrons Support (support@opentrons.com). They will work to design custom labware definitions based on your requirements. This is a fee-based service.

JSON labware schema

A JSON file is the blueprint for Opentrons standard and custom labware. This file contains and organizes labware data according to the design specifications set by the default schema.

A schema is a framework for organizing data. It sets the rules about what information is required or optional and how it’s organized in the JSON file. If you’re interested, take a moment to review our labware schemas. For an actual example, see the definition for the Opentrons 96 PCR Adapter. The following table defines the items in the Opentrons labware schema.

Property Data type Definition
schemaVersion Number Schema version used by a labware. The current version is 2.
version Integer An incrementing integer that identifies the labware version. Minimum version is 1.
namespace String See safeString in the JSON definitions section below.
metadata Object Properties used for search and display. Accepts only:
  • displayName (String): An easy-to-remember labware name.
  • displayCategory: Labels used in the UI to categorize labware. See displayCategory in the JSON definitions section below.
  • displayVolumeUnits (String): Labels used in the UI to indicate volume. Must be µL, mL, or L.
brand Object Information about the labware manufacturer or those products the labware is compatible with.
parameters Object Internal parameters that describe labware characteristics. Accepts only:
  • format (String): Determines labware compatibility with multichannel pipettes. Must be one of 96Standard, 384Standard, trough, irregular, or trash.
  • quirks (Array): Strings describing labware behavior. See the Opentrons 96 Deep Well Adapter definition.
  • isTiprack (Boolean): Indicates if labware is a tip rack (true) or not (false).
  • tipLength (Number): Required if labware is a tip rack. Specifies tip length (in mm), from top to bottom, as indicated in technical drawings or as measured with calipers.
  • tipoverlap (Number): Required if labware is a tip rack. Specifies how far tips on a tip rack are expected to overlap with the pipette's nozzle. Defined as tip length minus the distance between the bottom of the pipette and the bottom of the tip. The robot's calibration process may fine-tune this estimate.
  • loadName: Name used to reference a labware definition (e.g., opentrons_96_tiprack_300ul).
  • isMagneticModuleCompatible (Boolean): Indicates if labware is compatible with the Magnetic Module.
  • magneticModuleEngageHeight: How far the Magnetic Module will move its magnets when used with this labware. See positiveNumber in the JSON definitions section below.
ordering Array An array that tracks how wells should be ordered on a piece of labware. See the Opentrons 96 PCR Adapter example.
cornerOffsetFromSlot Object Used for labware that spans multiple deck slots. Offset is the distance from the left-front-bottom corner of the slot to the left-front-bottom corner of the labware bounding box. Accepts only:
  • x (number)
  • y (number)
  • z (number)
For labware that does not span multiple slots, these values should be zero. See positiveNumber in the JSON definitions section below.
dimensions Object Outer dimensions (in mm) of a piece of labware. Accepts only:
  • xDimension (length)
  • yDimension (width)
  • zDimension (height)
See the Opentrons 96 PCR Adapter example.
wells Object An unordered object of well objects, including position and dimensions.
Each well object's key is the well's coordinates, which must be an uppercase letter followed by a number, e.g., A1, B1, H12.
Each well object accepts the following properties:
  • depth (Number): The distance (in mm) between the top and bottom of the well. For tip racks, depth is ignored in favor of tipLength, but the values should match.
  • x (Number): Location of the center-bottom of a well in reference to the left of the labware.
  • y (Number): Location of the center-bottom of a well in reference to the front of the labware.
  • z (Number): Location of the center-bottom of a well in reference to the bottom of the labware.
  • totalLiquidVolume (Number): Total well, tube, or tip volume in µL.
  • xDimension (Number): Length of a rectangular well.
  • yDimension (Number): Width of a rectangular well.
  • diameter (Number): Diameter of a circular well.
  • shape (String): Either rectangular or circular.
    If rectangular, specify xDimension and yDimension.
    If circular, specify diameter.
For a circular well example, see the Opentrons 96 PCR Adapter. For a rectangular well example, see the NEST 96 Deep Well Plate 2mL.
For dimension, depth, and volume, see positiveNumber in the JSON definitions section below.
groups Array Logical well groupings for metadata and display purposes. Changes in groups do not affect protocol execution. Each item in the array accepts:
  • wells (Array): An array of wells (e.g., ["A1", "B1", "C1"]) that share the same metadata. Array elements are strings.
  • metadata (Object): Metadata specific to a grid of wells. Accepts only:
    • displayName (String): Human-readable name for the well group.
    • displayCategory: Labels used to categorize well groups. See displayCategory in the JSON definitions section below.
    • wellBottomShape (String): Bottom shape of a well. Available shapes are flat, u, or v only.
  • brand: Brand information for the well group. See brandData in the JSON definitions section below.
allowedRoles Array Defines an item's role or purpose. If the allowedRoles field is missing from a definition, an item is treated as labware. Possible array items are only the following strings:
  • labware (standard labware items)
  • adapter (items designed to hold labware)
  • fixture (items that are affixed to the deck)
  • maintenance (items not used in normal protocol runs)
stackingOffsetWithLabware Object For labware that can stack on top of another piece of labware. Used to determine z-height (labware z height + adapter z height - overlap). See coordinates in the JSON definitions section below.
stackingOffsetWithModule Object For labware that can stack on top of a module. Used to determine z-height (module labware offset z + labware z - overlap). See coordinates in the JSON definitions section below.
gripperOffsets Object Offsets added when calculating the coordinates the gripper should go to when picking up or dropping other labware on this labware. Includes a default object that includes two properties:
  • pickUpOffset: Offset added to calculate the pick-up coordinates of labware placed on this labware.
  • dropOffset: Offset added to calculate the drop-off coordinates of labware placed on this labware.
See coordinates in the JSON definitions section below.
gripForce Number Measured in newtons, this is the force which the gripper uses to grasp labware. Recommended values are between 5 and 16.
gripHeightFromLabwareBottom Number Recommended z-axis height, from the labware bottom to the center of the gripper pads.

JSON labware definitions

Property Data type Definition
positiveNumber Number Minimum: 0.
brandData Object Information about branded items. Accepts only:
  • brand (String): Brand/manufacturer's name.
  • brandId (Array): OEM part numbers or IDs.
  • links (Array): Manufacturer URLs. Array items are strings.
displayCategory String Must be one of:
  • tipRack
  • tubeRack
  • reservoir
  • trash
  • wellPlate
  • aluminumBlock
  • adapter
  • other
  • lid
safeString String A string safe to use for load names and namespaces. Lowercase letters, numerals, periods, and underscores only.
coordinates Object Coordinates that specify a distance or position along the x-, y-, and z-axes. Accepts only:
  • x (number)
  • y (number)
  • z (number)