Skip to content

System Specifications

General specifications

Specification Notes
Dimensions 66 cm x 57 cm x 63 cm (≈ 26" x 22.5" x 25")
Weight 48 kg (105 lbs)
Deck slots
  • 11 ANSI/SLAS-compliant slots in the working area.
  • 1 slot for a removable trash bin.
Wi-Fi 2.4 GHz IEEE 802.11b/g/n
Ethernet 100 Mbps
USB USB 2.0
Camera 2 megapixels
Power supply The OT-2 uses an external Mean Well GST220A series power supply. Refer to the manufacturer's data sheet (linked above) for detailed specifications.
Power consumption

Total power consumption depends on the state of your robot and the specific requirements of a protocol. For example:

  • When idle, the OT-2 can use 90 to 120 W.
  • When active, the OT-2 can use 100 to 180 W.

Exact power consumption is determined by the amount of movement executed during a protocol or the amount of time the robot spends idle. Power consumption may also be affected by the status of the lights on the OT-2 and the number of attached pipettes.

Frame composition Steel and aluminum.
Window composition Polycarbonate plastic.
Ventilation requirements Leave at least 20 cm (8") around the OT-2 to help ensure adequate ventilation.
Connected PC requirements

The Opentrons App runs on:

  • Windows 10 or later
  • macOS 10.16 or later
  • Ubuntu 20.04 or later