Unboxing

Congratulations! Your Opentrons Flex has arrived and you've prepared a space for it in your lab. Let's open that monster crate, remove the robot, and prepare it for operation. The information in this section provides a parts list and instructions that walk you through the steps required to get the Flex unboxed, set up, and ready for use. We've divided the setup procedure into three parts:

  • Part 1 covers disassembling the crate.

  • Part 2 covers detaching the Flex from the crate and moving it to a final assembly location.

  • Part 3 covers final assembly and powering on the robot for the first time.

Effort and time required

You'll want to ask a lab partner to assist with the unboxing, lifting, moving, and assembly process. You'll need to budget about 30 minutes to an hour for this effort.

Note

The Flex requires two people to lift it properly. Also, lifting and carrying the Flex by its handles is the best way to move the robot.

Crate and packing material

Unpacking a Flex gives you an awesome robot, but you're also left with several large crate panels along with assorted shipping components and padding. While you could discard this material, we encourage you to keep these items if storage space is available. The packaging is reusable, which helps prepare your Flex for shipping if you ever need to send it somewhere else (e.g., to a conference or a new facility) in the future.

Product elements

The Flex ships with the components listed below. Pipettes, the gripper, and modules come in separate packaging from the main Flex crate, even if you purchased them together as a workstation.

Opentrons Flex robot

(1) Opentrons Flex robot

USB cable

(1) USB cable

Ethernet cable

(1) Ethernet cable

Power cable

(1) Power cable

L-keys

(5) L-keys

(12 mm hex, 1.5 mm hex,
2.5 mm hex, 3 mm hex,
T10 Torx)

Emergency Stop Pendant

(1) Emergency Stop Pendant

Deck slot with labware clips

(1) Deck slot with labware clips

Spare labware clips

(4) Spare labware clips

Pipette calibration probe

(1) Pipette calibration probe

Carrying handles and caps

(4) Carrying handles and caps

Top window panel

(1) Top window panel

Side window panels

(4) Side window panels

2.5 mm hex screwdriver

(1) 2.5 mm hex screwdriver

19 mm wrench

(1) 19 mm wrench

Window screws

(16 + spares) Window screws

(M4x8 mm flat head)

Spare deck slot screws

(10) Spare deck slot screws

(M4x10 mm socket head)

Spare deck clip screws

(12) Spare deck clip screws

(M3x6 mm socket head)

Part 1: Remove the crate

Opentrons ships your Flex in a sturdy plywood crate. The shipping crate uses hook and latch clamps to secure the top, side, and bottom panels together. Using latches, instead of nails or screws, means you won't need a crowbar (or a lot of force) to disassemble the crate, and you can reassemble it later, if needed.

Note

Crate edges can get roughed up during shipping. You may want to use work gloves to protect your hands from wood splinters.

To release the latches, flip the latch tab up and turn it to the left (counterclockwise). This action moves the clamp arm out of its corresponding retaining bracket. You can then flip the latch arm away from the crate.

  1. Unlock the eight latches holding the top to the sides.

    Location of the eight Flex crate latches, with inset showing to turn them counterclockwise.

  2. Remove the top panel after releasing the latches.

  3. Cut open the blue shipping bag, remove these items from the padding, and set them aside:

    • User Kit

    • Power, Ethernet, and USB cables

    • Emergency Stop Pendant

    Exploded view of the locations of the user kit, cables, and Emergency Stop Pendant in the top foam padding.

  4. Remove the top piece of foam padding. The padding protects the installed top window panel.

  5. Unlock the remaining 16 latches holding the side panels to each other and the base of the crate.

  6. Remove the side panels and set them aside.

    Exploded view of the crate side panels.

Part 2: Release the Flex

After completing the steps in Part 1, you should now see a robot that's in a protective bag and attached to orange steel mounting components. The bag encloses the robot and protects it from the outside environment. Steel brackets secure the robot to the bottom of the crate. Two shipping frames support the robot, distributing its weight evenly, and keeping it rigid so it doesn't warp during shipping.

Continue to unpack the Flex and get it off the crate base.

  1. Using the 19 mm wrench from the User Kit, unbolt the brackets from the crate bottom. You can discard the brackets, or save them for future use.

    Location of the four crate bottom brackets, with inset showing how to remove them with the 19 mm wrench.

  2. Pull or roll the shipping bag all the way down to expose the entire robot.

    Shipping bag being removed from the Flex robot.

  3. With help from your lab partner, grab the handholds in the orange shipping frames on either side of the robot's base, lift the Flex off the crate base, and set it down on the floor. Save or discard the crate base and shipping frame.

    Location of handholds for lifting the Flex robot, with icon showing two people required to safely lift it.

  4. Using the 12 mm hex L-key from the User Kit, remove the four bolts holding the shipping frames to the Flex. Save or discard the frames and bolts.

    Location of the four shipping frame bolts, with inset showing how to loosen them by turning counterclockwise with the hex key.

  5. Remove the four aluminum handles from the User Kit. Screw the handles into the same locations that held the 12 mm shipping frame bolts.

    Location of the four handles with arrow showing clockwise attachment motion.

  6. With help from your lab partner, lift the Flex by its carrying handles and move it to a workbench for final assembly.

    Flex robot on top of a workbench.

Part 3: Final assembly and power on

After moving the Flex to a temporary work area, or its permanent home, it's time to put the finishing touches on your new robot.

  1. If you have moved the robot to its final, working location, remove the carrying handles and replace them with the finishing caps. The caps close the handle openings in the frame and give the robot a clean appearance. Return the handles to the User Kit for storage.

    Location of finishing caps, with inset showing clockwise attachment with the hex key.

  2. Using the 2.5 mm screwdriver from the User Kit, remove the locking screws from the gantry. These screws prevent the gantry from moving while in transit. The gantry locking screws are located:

    • On the left side rail near the front of the robot.

    • Underneath the vertical gantry arm.

    • On the right side rail near the front of the robot in an orange bracket. There are two screws here.

    Flex robot with three insets, each showing a locking screw location.

    The gantry moves easily by hand after removing all the shipping screws.

  3. Cut and remove the two rubber bands that hold the trash bin in place during shipping.

  4. Attach the power cord to Flex and plug it into the wall outlet. Make sure the deck area is free of obstructions. Flip the power switch on the back left of the robot. Once powered on, the gantry moves to its home location and the touchscreen displays additional configuration instructions.

    Location and orientation of the power plug on the back of the Flex robot.

Now that your Flex is out of the box and ready to go, continue to the First Run section.