Safety Information
Opentrons recommends that you follow the safe use specifications listed in this section and throughout this manual.
Safe Use Specifications¶
The Absorbance Plate Reader is designed for use with the Flex liquid handling robot. It cannot be used with an OT-2 robot or as standalone equipment. Follow the installation instructions to ensure proper and safe operation.
Warning
Always use the Opentrons Flex Gripper to move the illumination unit onto or off of the detection unit. Do not move the illumination unit by hand.
The Absorbance Plate Reader should only be used indoors in an environment with stable ambient conditions. The following table lists and defines the environmental operating conditions for recommended use, transportation, and storage of your plate reader.
Note
The module should not be powered on or used in settings that fall outside of the recommended operating conditions.
| Environmental Conditions | Recommended | Transportation | Storage | 
|---|---|---|---|
| Ambient Temperatures | +5 to +40 °C | -10 to 50 °C | +5 to +40 °C | 
| Relative Humidity | Up to 70% | Up to 70% | Up to 70% | 
The following table lists and defines standards for recommended use, transportation, and storage.
| Operating Conditions | Description | 
|---|---|
| Recommended | Opentrons has validated the performance of the Absorbance Plate Reader in the conditions recommended for system operation. Operating the module in these conditions provides optimal results. | 
| Transportation and storage | Storage and transportation conditions only apply when the device is completely disconnected from power and other equipment. | 
Instrument Safety Warnings¶
| Symbol | Description | 
|---|---|
| CAUTIONS: 
 |