Modules and fixtures

The second step in building your protocol is to add modules or fixtures to the deck. Click an open slot to add a module, like the Temperature Module, or a fixture like the trash bin.

Configure deck hardware

Click an open Flex deck slot to add compatible modules and fixtures.

Protocol Designer protocols can control Opentrons modules on the Flex or OT-2, as shown below. Flex lets you use multiple modules of the same type, except for the Thermocycler.

Module Opentrons Flex Opentrons OT-2
Thermocycler Module
  • Only 1 module
  • GEN2 only
  • Slots A1 and B1
  • Only 1 module
  • GEN1 or GEN2
  • Slots 7 and 10
Heater-Shaker Module
  • Up to 7 modules
  • GEN1 only
  • Slots A1–D1 or A3–D3
  • Only 1 module
  • GEN1 only
  • Slot 1, 3, 4, 6, 7, or 10
Absorbance Plate Reader Module
  • 1 module only
  • GEN1 only
  • Slots A3–D3
Not compatible
Magnetic Block
  • Up to 11 modules
  • GEN1 only
  • Slots A1–D3
Not compatible
Magnetic Module Not compatible
  • 1 module only
  • GEN1 or GEN2
  • Slot 1, 3, 4, 6, 7, 9, or 10
Temperature Module
  • Up to 7 modules
  • GEN2 only
  • Slots A1–D1 or A3–D3
  • 1 module only
  • GEN1 or GEN2 only
  • Slot 1, 3, 4, 6, 7, 9, or 10

On the Flex, you can also customize fixtures, including the waste chute, trash bin, and staging areas, to optimize deck space.

Click Deck hardware in the left sidebar, then click on a compatible slot to add a fixture.

Fixture Opentrons Flex Opentrons OT-2
Trash Bin
  • Only 1 trash bin
  • Slots A1-D1 or A3-D3
  • Only 1 trash bin
  • Slot 12
Waste Chute
  • Only 1 waste chute
  • Slot D3
Not compatible
Staging area
  • Up to 4 staging areas
  • Slots A4-D4
Not compatible

You can create additional space on the Flex deck by adding as many as four staging areas, creating slots A4-D4.