UV Specifications
The HEPA/UV Module holds two compact fluorescent UV bulbs. When on, the bulbs emit UV-C at the 254 nm wavelength. At this wavelength, UV-C disinfects by killing or damaging the genetic material found in various microorganisms. After a 15-minute exposure cycle, the ultraviolet light produced is sufficient to achieve log-4 (99.99%) inactivation of commonly targeted microorganisms within the enclosure.
Safety features¶
The HEPA/UV Module produces high-intensity ultraviolet light (UV-C). The Flex and the HEPA/UV Module come with features that protect you from UV-C exposure and prevent it from operating in an unsafe manner.
| Safety feature | Description | 
|---|---|
| Panels | The robot’s polycarbonate door and side panels block UV-spectrum light to below a level which represents an exposure risk. Never operate the UV lights with the robot’s door or side panels removed. | 
| Door switch | Flex uses a mechanical switch to tell if the front door is open or closed. The UV lights only work when the front door is closed. Opening the door deactivates/disables the UV lights. This safety interlock works even when Flex is turned off. | 
| Attachment sensor | The Flex and HEPA/UV Module each have a built-in sensor to detect if the module is attached properly. The sensors deactivate/disable the UV lights if the module is not mounted on the robot, removed while in operation, or misaligned. | 
Warning
- Direct exposure to UV-C light can damage your eyes and exposed skin. Always take care to protect your eyes and skin from exposure while the UV lights are on.
- Properly attach a fixture or deck plate cover in every deck slot before turning the UV lights on. Never operate the UV lights with empty deck cutouts.
Bulb data¶
| Specification | Description | 
|---|---|
| Bulb type | Actinic fluorescent light bulb | 
| Bulb base | 2G11 (4 pin, single cap) | 
| Useful life | 9,000 hours | 
| UV risk classification | 
 | 
| UV end-of-life deprecation | 20% | 
| Mercury (Hg) content | 4.4 mg | 
| Power consumption | 18 W | 
| Current (nominal) | 0.375 A | 
| Voltage (nominal) | 58 V |