Thermocycler

Additional Documentation
For complete instructions on module installation and use, see the Thermocycler Module Instruction Manual.
Thermocycler features¶
The Opentrons Thermocycler Module is a fully automated on-deck thermocycler designed for hands-free PCR in a 96-well plate format. It is compatible with the Flex Gripper, other deck-mounted hardware, and is fully supported in the Opentrons App and Python API. When used with a reusable rubber seal or single-use PCR plate lid, the module's heated lid provides a tight seal that helps ensure efficient sample heating and minimizes evaporation, crucial for reliable and repeatable experimental results.
Heating and cooling¶
The Thermocycler's block can heat and cool, and its lid can heat, with the following temperature profile:
- 
Thermal block temperature range: 4–99 °C 
- 
Thermal block maximum heating ramp rate: 4.25 °C/s from ambient to 95 °C 
- 
Thermal block maximum cooling ramp rate: 2.0 °C/s from 95 °C to ambient 
- 
Lid temperature range: 37–110 °C 
- 
Lid temperature accuracy: ±1 °C 
The automated lid can be opened or closed as needed during protocol execution.
Thermocycler profiles¶
The Thermocycler can execute profiles: automatically cycling through a sequence of block temperatures to perform heat-sensitive reactions.
Software control¶
The Thermocycler is fully programmable in Protocol Designer and the Python Protocol API.
Outside of protocols, the Opentrons App can display the current status of the Thermocycler and can directly control the block temperature, lid temperature, and lid position.
Thermocycler lids and seals¶
The Thermocycler is compatible with two different plate seals to help protect your samples: the single-use Opentrons Tough PCR Auto-sealing Lid and reuseable rubber automation seals.
Opentrons Tough PCR lids¶
The Opentrons Tough PCR Auto-sealing lids are single-use lids that prevent cross-contamination and evaporation during Thermocycler incubation periods. They are gripper-compatible and can be stacked directly on the deck or placed in a specialized deck riser.

These lids are manufactured in cleanroom facilities that meet ISO 14644-1 Class 8 standards. They also have a "Sterile-R" designation, which means they're irradiated to a sterility assurance level (SAL) of 10-6. Each box contains documentation certifying the lids are free of ATP, pyrogens, DNA, DNase, RNase, and PCR inhibitors.
Reuseable seals¶
These are adhesive-backed ethylene propylene diene monomer (EPDM) seals you manually apply to the Thermocycler lid. Each seal can be reused up to 20 times.

Unlike the Opentrons Tough PCR lids, reuseable seals are not sterile. They must be cleaned and sanitized before each use.
Warning
Do not use the Opentrons Tough Auto-sealing PCR Lid and a rubber automation seal on the Thermocycler at the same time. This combination prevents the module's lid from closing properly, which can cause temperature control problems and mechanical damage. Always remove the rubber seal before running protocols that use the disposable PCR lid.

Thermocycler specifications¶
| Specification | Details | 
|---|---|
| Dimensions (lid open) | 244.95 × 172 × 310.1 mm (L/W/H) | 
| Dimensions (lid closed) | 244.95 × 172 × 170.35 mm (L/W/H) | 
| Weight (including rear duct) | 8.4 kg | 
| Power adapter voltage | 100–240 V at 50/60 Hz | 
| Power adapter current | 8.5–5 A | 
| Overvoltage | Category II | 
| Environmental conditions | Indoor use only | 
| Ambient temperature | 20–25 °C (ideal); 2–40 °C (acceptable) | 
| Relative humidity | 30–80%, non-condensing | 
| Altitude | Up to 2000 m above sea level | 
| Ventilation requirements | At least 20 cm / 8 in between the unit and a wall |