Magnetic Block

Magnetic Block features¶
The Opentrons Magnetic Block GEN1 is a magnetic 96-well plate holder. Magnetic blocks are used in protocols that rely on magnetism to pull particles out of suspension and retain them in well plates during wash, rinse, or other elution procedures. For example, automated NGS preparation; purifying genomic and mitochondrial DNA, RNA, or proteins; and other extraction procedures are all use cases that can involve magnetic blocks.
Magnetic components¶
The Magnetic Block is unpowered, does not contain any electronic components, and does not move magnetic beads up or down in solution. The wells consist of 96 high-strength neodymium ring magnets fixed to a spring-loaded bed, which helps maintain tolerances between the block and pipettes while running automated protocols.
Software control¶
The Magnetic Block GEN1 is fully programmable in Protocol Designer and the Python Protocol API.
Outside of protocols, however, the touchscreen and the Opentrons App are not aware of and cannot display the current status of the Magnetic Block GEN1. This is an unpowered module. It does not contain electronic or mechanical components that can communicate with the Flex robot. You "control" the Magnetic Block via protocols that use the Opentrons Flex Gripper to add and remove labware from this module.
Magnetic Block specifications¶
| Specification | Details |
|---|---|
| Dimensions | 136 × 94 × 45 mm (L/W/H) |
| Weight | 1.13 kg |
| Module power | None, module is unpowered |
| Magnet grade | N52 neodymium |
| Environmental conditions | Indoor use only |
| Ambient temperature | 20–25 °C |
| Relative humidity | 30–80%, non-condensing |
| Altitude | Up to 2000 m above sea level |
| Pollution degree | 2 |